FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Human Plasma TMP & Thiamine by Gemini-NX 3u C18 100x3mm with Impact Plate_2.5nmol vs Blank
21880
Separation Mode: Reversed Phase
Reversed Phase
NX-C18
HPLC
Vitamin B Complex
Clinical

Applications

Human Plasma TMP & Thiamine by Gemini-NX 3u C18 100x3mm with Impact Plate_2.5nmol vs Blank
Analytes
1TMP
2Thiamine
Compound Name
Trimethoprim
CID
5578
Molecular Formula
C14H18N4O3
Molecular Weight
290
No. Hydrogen Bond Acceptors
7
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 21880)
Column:
Brand Name:
Part No:
Phase Name:
NX-C18
Elution Type:
Gradient
Mobile Phase:
A: 25nM Na2HPO4, 10% methanol
B: 25mM Na2HPO4, 70% methanol
Gradient Profile:
Step No.Time(min)%A%B
10397
20.252575
30.752575
433565
541000
651000
75.1397
88397
Flow Rate:
0.75 mL/min
Temperature:
22°C
Detection:
Fluorescence (FLUOR) @ Ex 375 nm (22°C)
Detector Info:
Shimadzu ELSD
Sample Preparation Details
Description:
Impact Protein Precipitation, 2mL Square Well Filter Plate, 2/Pk
Part Number:
Evaporation:
Evaporate to dryness

Order Items Used in This Application

Similar Applications
search
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase