For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Bath salts and Piperazines using Kinetex 2.6u Biphenyl 50x3.0mm
22152
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Serotonin Agonists
Pharmaceutical/Biopharmaceutical
Clinical
Forensics/Toxicology
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Bath salts and Piperazines using Kinetex 2.6u Biphenyl 50x3.0mm
Analytes
1BZP
2Methylone
3Mephedrone
4mCPP
5TFMPP
6MDPV
Compound Name
1-Benzylpiperazine
CID
75994
Molecular Formula
C11H16N2
Molecular Weight
176
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 22152)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Biphenyl
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% formic acid
B: Methanol with 0.1% formic acid
Gradient Profile:
Step No.Time(min)%A%B
109010
22.5595
33595
43.19010
54.59010
Flow Rate:
0.6 mL/min
Temperature:
40°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)