FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Cefaclor forced degradation on Kinetex 2.6 µm EVO C18
22988
Separation Mode: Reversed Phase
Reversed Phase
EVO C18
HPLC
Anti-Bacterial Agents
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

Cefaclor forced degradation on Kinetex 2.6 µm EVO C18
Analytes
1Cefaclor
Compound Name
Cefaclor
CID
51039
Molecular Formula
C15H14ClN3O4S
Molecular Weight
367
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
3
Details
LC Conditions (App ID: 22988)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
EVO C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% H3PO4 in Water
B: 0.1% H3PO4 in ACN
Gradient Profile:
Step No.Time(min)%A%B
10955
266040
Flow Rate:
1.75 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
Forced degradation protocol: Reconstitute 1 mg of cefaclor neat standard with 1.0 mL of 1 M HCl. Heat the solution at 60 ºC for 6 hours then neutralize with 1.0 mL of 1 M NaOH. Refrigerate until ready to inject.