FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Drugs of Abuse by LCMS using Kinetex 1.7u EVO C18 50x2.1mm
23444
Separation Mode: Reversed Phase
Reversed Phase
EVO C18
HPLC
Analgesics, Opioid
Pharmaceutical/Biopharmaceutical
Forensics/Toxicology
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Drugs of Abuse by LCMS using Kinetex 1.7u EVO C18 50x2.1mm
Analytes
1BZE
2Morphine
3Hydromorphone
4Amphetamine
5MDA
6Oxymorphone
7MDMA
8Methamphetamine
9Codeine
106-MAM
11MDEA
12Hydrocodone
13Oxazepam
14Lorazepam
15Alprazolam
16Oxycodone
17Clonazepam
18Norbuprenorphine
19Temazepam
20Midazolam
21Diazepam
22Flurazepam
23Methadone
24PCP
25Buprenorphine
Compound Name
Benzoylecgonine
CID
448223
Molecular Formula
C16H19NO4
Molecular Weight
289
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 23444)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
EVO C18
Elution Type:
Gradient
Mobile Phase:
A: 10mM Ammonium Bicarbonate pH 10
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
10955
24595
35595
45.1955
Flow Rate:
0.4 mL/min
Temperature:
40°C
Detection:
Tandem Mass Spec (MS-MS) @ 254 nm (22°C)
Detector Info:
API 4000

Order Items Used in This Application

Similar Applications
search
Kinetex 1.7 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm Biphenyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna Omega 5 µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Luna Omega 3 µm Polar C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna Omega 1.6µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Luna Omega 1.6µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Luna Omega 1.6µm C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase