FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
ASMS 2016 Poster Figure 12_ Matrix of Water vs Whole Blood - Pyridoxal 5-phosphate (PLP)
23557
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC

Applications

ASMS 2016 Poster Figure 12_ Matrix of Water vs Whole Blood - Pyridoxal 5-phosphate (PLP)
Analytes
1Pyridoxal 5-phosphate (PLP)

Analyte not found

Details
LC Conditions (App ID: 23557)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 10 mM NH4HCO3 pH 8.8
B: MeOH
Gradient Profile:
Step No.Time(min)%A%B
10.011000
21.51000
31.84060
454060
55.51000
681000
Flow Rate:
500 µL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (600°C)
Detector Info:
SCIEX Triple Quad™ 4500 System