FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Taxanes on Kinetex 5µm F5
24258
Separation Mode: Reversed Phase
Reversed Phase
F5
HPLC
Antineoplastic Agents, Phytogenic

Applications

Taxanes on Kinetex 5µm F5
Analytes
17-Epi 10-Desacetyl Paclitaxel
210-deacetylbaccation III
3Baccatin III
47-epi-taxol
5Taxol (Paclitaxel)

Analyte not found

Details
LC Conditions (App ID: 24258)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
F5
Elution Type:
Gradient
Mobile Phase:
A: Water + 0.1 % Formic Acid
B: Acetonitrile + 0.1 % Formic Acid
Gradient Profile:
Step No.Time(min)%A%B
107525
2203070
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 230 nm (22°C)