FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
TCAs on Kinetex 2.6u Polar C18, 50x2.1mm (100 ng/mL)
24276
Separation Mode: Reversed Phase
Reversed Phase
Polar C18
HPLC
Antidepressive Agents, Tricyclic

Applications

TCAs on Kinetex 2.6u Polar C18, 50x2.1mm (100 ng/mL)
Analytes
1Nortriptyline
2Protriptyline
3Desmethyldoxepin
4Desipramine
5Amitriptyline
6Doxepin
7Imipramine
8N-Desmethylclomipramine
9Clomipramine
Compound Name
Nortriptyline
CID
4543
Molecular Formula
C19H21N
Molecular Weight
263
No. Hydrogen Bond Acceptors
1
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 24276)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Polar C18
Elution Type:
Gradient
Mobile Phase:
A: 10 mM ammonium formate in water
B: 0.1% formic acid in methanol
Gradient Profile:
Step No.Time(min)%A%B
105545
223565
32.5595
43595
53.15545
64.55545
Flow Rate:
0.5 mL/min
Temperature:
40°C
Detection:
Tandem Mass Spec (MS-MS) (600°C)
Detector Info:
SCIEX Triple Quad™ 4500 System