FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Naproxen Oral Suspension - USP Acceptable Modified Method
3283
Separation Mode: Reversed Phase
Reversed Phase
USP L01
C18(2)
HPLC
Anti-Inflammatory Agents, Non-Steroidal
Clinical
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

Naproxen Oral Suspension - USP Acceptable Modified Method
Analytes
1Naproxen
2Ethyl paraben
Compound Name
Naproxen
CID
156391
Molecular Formula
C14H14O3
Molecular Weight
230
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
1
Details
Official Method: USP L01
LC Conditions (App ID: 3283)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Isocratic
Mobile Phase:
A: Methanol
B: Water with 30mM Sodium acetate pH 5.8
Gradient Profile:
Step No.Time(min)%A%B
105050
Flow Rate:
2 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C8 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C8 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C8 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C8 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Polar-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase