For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Polycyclic Aromatic Hydrocarbons (PAHs) in Water byZebron ZB-5MS
19651
Separation Mode: Standard
Standard
ZB-5MS
GC
Carcinogens
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

Polycyclic Aromatic Hydrocarbons (PAHs) in Water byZebron ZB-5MS
Analytes
1Naphthalene-d8
2Naphthalene
32-Methylnaphthalene
4Acenaphthalene
5Acenaphthene
6Fluorene
7Anthracene
8Phenanthrene
9Fluoranthene
10Pyrene-d10
11Pyrene
12Benzo[a]anthracene
13Chrysene
14Benzo[b]fluroanthene
15Benzo[k]fluroanthene
16Benzo[a]pyrene-d12
17Benzo[a]pyrene
18Indeno[1,2,3-cd]pyrene
19Dibenz[a,h]anthracene
20Benzo[g,h,i]perylene
Compound Name
Naphthalene-D8
CID
92148
Molecular Formula
C10H8
Molecular Weight
136
Details
GC Conditions (App ID: 19651)
Column:
Part No:
Phase Name:
ZB-5MS
Injection:
Split 15:1 @ 310°C, 1.0 μL
Oven Program:
140C to 240C @ 15C/min to 275C @ 4C/min to 320C @ 10C/min for 5min
Carrier Gas:
Constant Flow Helium, 1.4 mL/min
Detection:
Mass Spectrometer (MS) (270°C)
Detector Info:
Within the 17 PAH's Mix Deuterated versions of 3 of the probes were added, surrogates (D8 Napthalene, D10 Pyrene) and D12-Benzo[a]pyrene for the ISS

Order Items Used in This Application

Similar Applications
search
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 20 m x 0.18 mm x 0.36 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.32 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard