FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Lisinopril Ph. Eur. System Suitability Reference Solution A on Kinetex 5 µm C18 250 x 4.6 mm on Agilent 1100 LC/UV
26295
Separation Mode: Reversed Phase
Reversed Phase
Ph.Eur. M.1120
C18
HPLC
Antihypertensive Agents
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Applications

Lisinopril Ph. Eur. System Suitability Reference Solution A on Kinetex 5 µm C18 250 x 4.6 mm on Agilent 1100 LC/UV
Analytes
1Impurity A
2Lisinopril
3Impurity E
Compound Name
Benzylalanine
CID
102530
Molecular Formula
C10H13NO2
Molecular Weight
179
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
2
Details
Official Method: Ph.Eur. M.1120
LC Conditions (App ID: 26295)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: Mix 3 volumes of acetonitrile R1 and 97 volumes of 3.12 g/L solution of sodium dihydrogen phosphate R previously adjusted to pH 3.8 with dilute phosphoric acid R
B: Mix 20.5 volumes of acetonitrile R1 and 79.5 volumes of 3.12 g/L solution of phosphoric acid R
Gradient Profile:
Step No.Time(min)%A%B
10100
22100
3370100
4620100
562.5100
666100
Flow Rate:
1.6 mL/min
Temperature:
50°C
System:
Agilent Technologies Agilent Technologies
Detection:
LC/UV (DAD, PDA) @ 210.0000000000 nm (nanometers)

Order Items Used in This Application

Similar Applications
search
Kinetex 5 µm XB-C18 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm C18 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm C18 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase