For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Polynuclear Aromatic Hydrocarbons (PAHs) on ZB-5MS
15560
Separation Mode: Standard
Standard
ZB-5MS
GC
Carcinogens
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

Polynuclear Aromatic Hydrocarbons (PAHs) on ZB-5MS
Analytes
1Naphthalene
22-Methylnaphthalene
3Acenaphthalene
4Acenaphthene
5Fluorene
6Phenanthrene
7Anthracene
8Fluoranthene
9Pyrene
10Benz[a]anthracene
11Chrysene
12Benzo[b]fluoranthene
13Benzo[k]fluoranthene
14Benzo[a]pyrene
15Indeno[1,2,3-cd]pyrene
16Dibenz[a,h]anthracene
17Benzo[g,h,i]perylene
Compound Name
Naphthalene
CID
931
Molecular Formula
C10H8
Molecular Weight
128
Details
GC Conditions (App ID: 15560)
Column:
Part No:
Phase Name:
ZB-5MS
Injection:
Split 15:1 @ 285°C, 1.0 μL
Oven Program:
130°C for 0.5min to 250°C at 25°C/min to 270°C at 6°C/min to 320°C at 25°C/min for 4 min
Carrier Gas:
Constant Flow Helium, 1.2 mL/min
Detection:
Mass Selective (MSD) (180°C)

Order Items Used in This Application

Similar Applications
search
Kinetex 3.5 µm PAH, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Kinetex 3.5 µm PAH, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Reversed Phase