For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 515.2: Chlorinated Acids on ZB-1701
15791
Separation Mode: Standard
Standard
EPA 515
ZB-1701
GC
Herbicides
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA Method 515.2: Chlorinated Acids on ZB-1701
Analytes
13,5-Dichlorobenzoic acid
22,4-Dichlorophenylacetic acid
3Dicamba
4Dichloroprop
54,4'-Dibromooctafluorobiphenyl
62,4-D
7Pentachlorophenol
8Silvex
92,4,5-T
102,4-DB
11Dinoseb
12Bentazon
13Dacthal
14Picloram
15Acifluorfen
Compound Name
Diobutil
CID
5811
Molecular Formula
C7H4Cl2O2
Molecular Weight
191
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 515
GC Conditions (App ID: 15791)
Column:
Part No:
Phase Name:
ZB-1701
Injection:
Splitless @ 200°C, 2.0 μL
Oven Program:
80°C to 180°C at 25°C/min to 220°C at 8°C/min to 260°C at 20°C/min for 3 min
Carrier Gas:
Constant Flow Helium, 1.2 mL/min
Detection:
Electron Capture (ECD) (280°C)
Sample Notes:
Analytes were methylated

Order Items Used in This Application

Similar Applications
search
Synergi 4 µm Hydro-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.32 mm x 0.50 µm, Ea
Application Specific