For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA 8141: Organophosphorous Pesticides on ZB-MR1
16181
Separation Mode: Application Specific
Application Specific
EPA 8141
ZB-MultiResidue-1
GC
Antinematodal Agents
Environmental
Pesticide / Insecticide
Environmental / Health

Applications

EPA 8141: Organophosphorous Pesticides on ZB-MR1
Analytes
1Dichlorvos
2Mevinphos
3Trichlorfon
4TEPP
5Demeton
6Thionazin
7Ethoprop
8Sulfotep
9Naled
10Dicrotophos
11Phorate
12Monocrotophos
13Demeton
14Terbufos
15Diazinon
16Dimethoate
17Fonofos
18Phosphamidon
19Disulfoton
20Dichlofenthion
21Phosphamidon
22Chlorpyrifos methyl
23Ronnel
24Aspon
25Methyl parathion
26Malathion
27Fenitrothion
28Chlorpyrifos
29Fenthion
30Trichloronate
31Parathion
32Merphos
33Chlorfenvinphos
34Crotoxyphos
35Tetrachlorvinphos
36Tokuthion
37Merphos oxide
38Ethion
39Fensulfothion
40contaminant
41Carbofenothion
42Famphur
43EPN
44Phosmet
45Leptophos
46Azinphos methyl
47Azinphos ethyl
48Coumaphos
Compound Name
Dichlorvos
CID
3039
Molecular Formula
C4H7Cl2O4P
Molecular Weight
220
No. Hydrogen Bond Acceptors
4
Details
Official Method: EPA 8141
GC Conditions (App ID: 16181)
Column:
Part No:
Phase Name:
ZB-MultiResidue-1
Injection:
On-Column @ 103°C, 1.0 μL
Oven Program:
100 °C for 0.5 min to 180 °C at 20 °C/min to 240 °C at 6 °C/min to 320 °C at 15 °C/min for 2 min
Carrier Gas:
Constant Flow Helium, 2.8 mL/min
Detection:
Flame Ionization (FID) (340°C)
Sample Notes:
Analytes are 2 ppm in Dichloromethane. Columns connected via a 5 m guard and 'Y' connector.

Order Items Used in This Application

Similar Applications
search
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
Application Specific
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
Application Specific
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-1701P, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-5MS, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard