FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Jupiter® 300 C4 - PEGylation Reaction Time Course (Carbonic Anhydrase)
16198
Separation Mode: Reversed Phase
Reversed Phase
C4
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Biochemical - Polymeric
Biochemical Compound / Nutrient

Applications

Jupiter® 300 C4 - PEGylation Reaction Time Course (Carbonic Anhydrase)
Analytes
1PEGylated Carbonic anhydrase

Analyte not found

Details
LC Conditions (App ID: 16198)
Column:
Brand Name:
Part No:
Phase Name:
C4
Elution Type:
Gradient
Mobile Phase:
A: 0.1% TFA in Water
B: 90% ACN 0.09% TFA in Water
Gradient Profile:
Step No.Time(min)%A%B
108020
2254555
Flow Rate:
1 mL/min
Temperature:
45°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 214 nm (22°C)
Sample Note:
Application Focus: To demonstrate the utility of using reversed phase chromatography with Jupiter 300 media for purifying PEGylated proteins. The addition of PEG groups to a protein complicates both the characterization and purification of such PEG/protein conjugates away from the "non-PEGylated" protein species. Protein PEGylation was performed using Methyl-PEO12-NHS Ester. A protein solution in PBS (pH 7.4) was reacted with an excess of PEGylation reagent dissolved in DMSO and incubated at 4C for up to two hours. The reaction was quenched with 50 mM Tris/1 % trifluoroacetic acid. In general, the PEGylation reaction concurrently occurs rapidly at several different protein sites in a fixed ratio. As the reaction continues, more heavily PEGylated (and later eluting) forms were observed. In every protein tested there was always more than one PEGylated protein peak observed by reversed phase HPLC; each seemingly ascribed to a different site of PEG modification. Unlike gel filtration which only separates PEGylated proteins by degree of modification, it appears that protein modification site influences the reversed phase separation of proteins much more than the degree of modification; such results show the utility of using reversed phase chromatography for purifying and analyzing PEGylated proteins.

Order Items Used in This Application

Similar Applications
search
Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Jupiter 5 µm C4 300 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase