FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
USP 467 Procedure A - Class 1 by GC/FID on ZB-624
16348
Separation Mode: Standard
Standard
USP <467>
ZB-624
GC
Pharmaceutical/Biopharmaceutical

Applications

USP 467 Procedure A - Class 1 by GC/FID on ZB-624
Analytes
11,1-Dichloroethene
21,1,1-Trichloroethane
3Carbon tetrachloride
4Benzene
51,2-Dichloroethane
Compound Name
Vinylidene Chloride
CID
6366
Molecular Formula
C2H2Cl2
Molecular Weight
96
Details
Official Method: USP <467>
GC Conditions (App ID: 16348)
Column:
Part No:
Phase Name:
ZB-624
Injection:
Headspace 5:1 @ 140°C, 1.0 μL
Oven Program:
40 °C for 20 min to 240 °C @ 10 °C/min for 20 min
Carrier Gas:
Constant Flow Helium, 35.0 cm/sec
Detection:
Flame Ionization (FID) (250°C)
Sample Notes:
Sample Prepared as per USP Guidelines