FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 8151A on ZB-35 (Carrier Gas: Helium)
16758
Separation Mode: Standard
Standard
EPA 8151
ZB-35
GC
Herbicides
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA Method 8151A on ZB-35 (Carrier Gas: Helium)
Analytes
1Dalapon
23,5-Dichlorobenzoic acid
34-Nitrophenol
4DCAA
5Dicamba
6MCPP
7MCPA
8DBOB
9Dichloroprop
102,4-D
11Pentachlorophenol
12contaminant
13Silvex
142,4,5-T
15Chloramben
16Dinoseb
172,4-DB
18Bentazone
19Picloram
20Dacthal
21Acifluorfen
Compound Name
Kenapon
CID
6418
Molecular Formula
C3H4Cl2O2
Molecular Weight
142
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 8151
GC Conditions (App ID: 16758)
Column:
Part No:
Phase Name:
ZB-35
Injection:
Splitless @ 250°C, 1.0 μL
Oven Program:
50 °C for 0.5 min to 100 °C @ 25 °C/min to 320 °C @ 12 °C/min for 2 min.
Carrier Gas:
Constant Flow Helium, 2.4 mL/min
Detection:
Electron Capture (ECD) (300°C)
Sample Notes:
Analyes at 0.5 ppm in Methanol

Order Items Used in This Application

Similar Applications
search
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-CLPesticides-2, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-CLPesticides-1, GC Cap. Column 30 m x 0.32 mm x 0.32 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-35HT, GC Cap. Column 30 m x 0.32 mm x 0.25 µm, Ea
High Temperature