For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Extraction of Testosterone from Human plasma using strata-X on Gemini C18
17473
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Androgens
Clinical
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Applications

Extraction of Testosterone from Human plasma using strata-X on Gemini C18
Analytes
1Testosterone
Compound Name
Testosterone
CID
6013
Molecular Formula
C19H28O2
Molecular Weight
288
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 17473)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Isocratic
Mobile Phase:
A: 0.1% formic acid in water
B: 0.1% formic acid in acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
113070
Flow Rate:
0.5 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Gemini C18 4 x 2.0mm, 10/Pk Part No.: AJ0-7596 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Kinetex 5 µm EVO C18 100 Å, LC Column 100 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Gemini 3 µm NX-C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Gemini 5 µm C18 110 Å, LC Column 50 x 2 mm, Ea
Reversed Phase
Luna 5 µm CN 100 Å, LC Column 150 x 4.6 mm, Ea
Normal Phase