For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Mianserin on Lux 5µm Cellulose-1
17477
Separation Mode: Chiral
Chiral
Cellulose-1
HPLC
Antidepressive Agents
Pharmaceutical/Biopharmaceutical
Therapeutic / Clinical
Therapeutic / Clinical
Chemical Compound

Applications

Mianserin on Lux 5µm Cellulose-1
Analytes
1Mianserin
2Mianserin
Compound Name
Mianserin
CID
4184
Molecular Formula
C18H20N2
Molecular Weight
264
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
2
Details
LC Conditions (App ID: 17477)
Column:
Brand Name:
Part No:
Phase Name:
Cellulose-1
Elution Type:
Isocratic
Mobile Phase:
A: 0.1 % DEA in CH3CN
B: 0.1 % DEA in IPA
Gradient Profile:
Step No.Time(min)%A%B
110955
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 220 nm (22°C)
Detector Info:
alpha=1.28 and Rs=2.61
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Lux Cellulose-1, 4 x 3.0mm, 10/Pk Part No.: AJ0-8403 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Lux 5 µm i-Amylose-1, LC Column 250 x 4.6 mm, Ea
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral
Lux 5 µm Amylose-1, LC Column 250 x 4.6 mm, EA
Chiral