FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Organochloropesticides (OCP) and Polychlorinated Biphenyls (PCB) in Milk and Milk Products by GC/ECD
17910
Separation Mode: Application Specific
Application Specific
ZB-MultiResidue-2
GC
Pesticides
Food & Beverage
Pesticide / Insecticide
Environmental / Health

Applications

Organochloropesticides (OCP) and Polychlorinated Biphenyls (PCB) in Milk and Milk Products by GC/ECD
Analytes
1Hexachlorobenzene
2alpha-HCH
3gamma-HCH
4beta-HCH
5PCB 28
6delta-HCH
7Aldrin
8PCB 52
9Heptachloro epoxide
102,4'-DDE (o,p-)
11PCB 101
124,4'-DDE (p,p-)
13Nitrofen
14Dieldrin
152,4'-DDD (o,p-)
16Endrin
174,4'-DDD (p,p-)
18Nitrofen
19PCB 153
20PCB 138
21PCB 180
22Mirex
Compound Name
Perchlorobenzene
CID
8370
Molecular Formula
C6Cl6
Molecular Weight
284
Details
GC Conditions (App ID: 17910)
Column:
Part No:
Phase Name:
ZB-MultiResidue-2
Injection:
Splitless @ 50°C, 5.0 μL
Oven Program:
100 °C for 1 min, with 35 °C/min to 165 °C and 22 min hold, with 20 °C/min to 190 °C, with 4 °C/min to 250°C and 1 min hold
Carrier Gas:
Constant Flow Hydrogen, 1.5 mL/min
System:
Agilent (HP) 6890
Detection:
Electron Capture (ECD) (300°C)
Sample Notes:
The application was kindly provided by the MLF, Dr.-Oskar-Farny-Institute in Wangen, Germany. Injector Gerstel-KAS3. Program: 50 °C with 12 °C/s to 300 °C, 3 min hold Solvent Venting Purge Time: 0.15 min, Splitless Time: 1 min, Equilibration Time 0.5 min, Initial Time: 0.15 min. Concentration: 25 ppb each for OCP + PCB, 100 ppb for Nitrofen