For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Cocaine and Metabolites on Luna HILIC 3u 100x2.0
17914
Narcotics
Forensics/Toxicology
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Applications

Cocaine and Metabolites on Luna HILIC 3u 100x2.0
Analytes
1Cocaine
2Norcocaine
3Ecgonine methyl ester
4Benzoylecgonine
5Ecgonine
Compound Name
Cocaine
CID
5760
Molecular Formula
C17H21NO4
Molecular Weight
303
No. Chiral Centers:
2
No. Hydrogen Bond Acceptors
5
Details
LC Conditions (App ID: 17914)
Elution Type:
Gradient
Mobile Phase:
A: 95/5 ACN/100mM NH4 Formate pH 3.2
B: 50/45/5 ACN/H2O/100mM NH4 Formate pH 3.2
Gradient Profile:
Step No.Time(min)%A%B
101000
21.71000
34.84555
46.54555
56.61000
612.61000
Flow Rate:
0.4 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
Mass Transitions Cocaine 304 to 181.9 Norcocaine 290.2 to 168.2 & 290.1 to 136.0 Ecgonine methyl ester 200.2 to 182.1 Benzoylecgonine 290.2 to 168.2 Ecgonine 186.2 to 168.1
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, HILIC 4 x 2.0mm, 10/Pk Part No.: AJ0-8328 - Holder Part No.: KJ0-4282
Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Zebron ZB-5MSplus, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, EA
Improved Performance
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea
Standard
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific
Zebron ZB-XLB, GC Cap. Column 30 m x 0.25 mm x 0.50 µm, Ea
Standard
Zebron ZB-1HT, GC Cap. Column 30 m x 0.25 mm x 0.10 µm, Ea
High Temperature
Zebron ZB-50, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard