FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Interferon Alpha intact & oxidized on Jupiter 3u C18 and Jupiter 5u C4
18074
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Clinical Research
Therapeutic / Clinical

Applications

Interferon Alpha intact & oxidized on Jupiter 3u C18 and Jupiter 5u C4
Analytes
1Intact & Oxidized Interferon

Analyte not found

Details
LC Conditions (App ID: 18074)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% TFA and 2% Acetonitrile in Water
B: 0.085% TFA, 90% Acetonitrile in Water
Gradient Profile:
Step No.Time(min)%A%B
108020
2102080
3151090
Flow Rate:
0.3 mL/min
Temperature:
25°C
Detection:
UV-Vis Abs.-Diode Array (PDA) @ 220 nm (25°C)
Sample Note:
Application Focus: Using Jupiter 300 media for development of intact biogeneric protein assays for oxidation. Physical and chemical degradation of therapeutic proteins is a critical problem that can occur during production, purification, and storage. Such modifications can affect protein immunogenicity leading to serious consequences if the protein is being used as a therapeutic. Chromatograms comparing native to oxidized alpha interferon were overlaid to visualize chromatographic differences; oxidation of intact Interferon-alpha was performed by treating with 3% hydrogen peroxide and 50 mM sodium citrate pH 4.4 at 45°C for 120 min. Overlaid chromatographs of the intact versus oxidized alpha interferon clearly show good selectivity between the two samples; oxidized interferon elutes earlier than the intact protein and has a dramatically tailing peak. While both the C4 and C18 phases both had good resolution, the 3µm C18 clearly showed better performance than all other phases tested. These results indicate that the Jupiter media is an excellent solution with the required selectivity for quantitating very low levels of oxidized protein in less than 15 minutes.