For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Expanded amphetamines from urine by Strata-XC on a ZB-Drug1
18482
Separation Mode: Application Specific
Application Specific
ZB-Drug-1
GC
Serotonin Agents
Forensics/Toxicology
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Applications

Expanded amphetamines from urine by Strata-XC on a ZB-Drug1
Analytes
1Amphetamine-d11
2Amphetamine
3Methamphetamine-d14
4Methamphetamine
5MDA-d5
6MDA
7MDMA-d5
8MDMA
9MDEA-d5
10MDEA
Compound Name
Levamfetamine
CID
3056364
Molecular Formula
C9H13N
Molecular Weight
146
No. Chiral Centers:
1
No. Hydrogen Bond Acceptors
1
No. Hydrogen Bond Donors
1
Details
GC Conditions (App ID: 18482)
Column:
Part No:
Phase Name:
ZB-Drug-1
Injection:
Split 10:1 @ 260°C, 1.0 μL
Oven Program:
70 °C to 130 °C @ 20 °C/min to 320 °C @ 30 °C/min for 0.67 min
Carrier Gas:
Constant Flow Helium, 0.7 mL/min
Detection:
Mass Selective (MSD) (320°C)
Sample Notes:
Analytes are 500ng/mL in urine derivatized with HFBA Sim ions: (Quant ion*) D11-Amphetamine (*244,128): Amphetamine (*240,118,117): D14-Methamphetamine (*261,213): Methamphetamine (*254,210,118): D5-MDA (*167,380,244): MDA (*162,375,240): D5-MDMA (*258,213,164): MDMA (*254,210,389): D5-MDEA (*273,241,408): MDEA (*268,240,403)

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm XB-C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C8 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Zebron ZB-Drug-1, GC Cap. Column 10 m x 0.18 mm x 0.18 µm, Ea
Application Specific