For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA 604 - Fast Separation of Phenols Mixture Using Kinetex 2.6µm C18 50 x 4.6 mm
18677
Separation Mode: Reversed Phase
Reversed Phase
EPA 604
C18
HPLC
Uncoupling Agents
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA 604 - Fast Separation of Phenols Mixture Using Kinetex 2.6µm C18 50 x 4.6 mm
Analytes
1Phenol
24-Nitrophenol
32-Chlorophenol
42-Nitrophenol
52,4-Dimethylphenol
62,4-Dinitrophenol
74-Chloro-3-methylphenol
82,4-Dichlorophenol
92-Methyl-4,6-dichlorophenol
102,4,6-Trichlorophenol
11Pentachlorophenol
Compound Name
Phenol
CID
996
Molecular Formula
C6H6O
Molecular Weight
94
No. Hydrogen Bond Acceptors
1
No. Hydrogen Bond Donors
1
Details
Official Method: EPA 604
LC Conditions (App ID: 18677)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% H3PO4
B: Acetonitrile with 0.1% H3PO4
Gradient Profile:
Step No.Time(min)%A%B
107525
21.5595
Flow Rate:
2.2 mL/min
Temperature:
0°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 214 nm (22°C)

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 20 m x 0.18 mm x 0.36 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.32 µm, Ea
Standard