For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Peptide Map of biogeneric Alpha Interferon (unreduced) on Kinetex 2.6µm C18 150 x 4.6mm ID
18767
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Pharmaceutical
Therapeutic / Clinical

Applications

Peptide Map of biogeneric Alpha Interferon (unreduced) on Kinetex 2.6µm C18 150 x 4.6mm ID
Analytes
1Human alpha Interferon

Analyte not found

Details
LC Conditions (App ID: 18767)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% TFA/2% ACN/Water
B: 0.09% TFA/98% ACN/Water
Gradient Profile:
Step No.Time(min)%A%B
10991
2404456
3414456
Flow Rate:
1 mL/min
Temperature:
0°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 214 nm (22°C)

Order Items Used in This Application

Similar Applications
search
No Data