For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 8081- Organochlorine Pesticides on ZB-35HT
18781
Separation Mode: High Temperature
High Temperature
EPA 8081A
ZB-35HT
GC
Insecticides
Environmental
Pesticide / Insecticide
Environmental / Health

Applications

EPA Method 8081- Organochlorine Pesticides on ZB-35HT
Analytes
1alpha-BHC
2gamma-BHC
3b-BHC
4Heptachlor
5delta-BHC
6Aldrin
7Heptachlor epoxide
8gamma-Chlordane
9alpha-Chlordane
10Endosulfan I
11DDE
12Dieldrin
13Endrin
14DDD
15Endosulfan II
16DDT
17Endrin aldehyde
18Endosulfan sulfate
19Methoxychlor
20Endrin ketone
Compound Name
Lindane
CID
727
Molecular Formula
C6H6Cl6
Molecular Weight
290
Details
Official Method: EPA 8081A
GC Conditions (App ID: 18781)
Column:
Part No:
Phase Name:
ZB-35HT
Injection:
Split 5:1 @ 350°C, 1.0 μL
Oven Program:
160 °C for 0.5 min to 320 °C @ 25 °C/min for 1.5 min
Carrier Gas:
Constant Flow Helium, 1.2 mL/min
Detection:
Electron Capture (ECD) (350°C)
Sample Notes:
Analytes are 50ppb in isooctane

Order Items Used in This Application

Similar Applications
search
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-35HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature