FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Ethyl Sulfate (25ng/mL) on Synergi Hydro-RP (m/z 124.9 > 87.0)
19781
Separation Mode: Reversed Phase
Reversed Phase
Hydro-RP
HPLC
Clinical

Applications

Ethyl Sulfate (25ng/mL) on Synergi Hydro-RP (m/z 124.9 > 87.0)
Analytes
1EtS (Ethyl sulfate)
Compound Name
Sulfovinic Acid
CID
6004
Molecular Formula
C2H6O4S
Molecular Weight
126
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 19781)
Column:
Brand Name:
Part No:
Phase Name:
Hydro-RP
Elution Type:
Gradient
Mobile Phase:
A: 10mM Ammonium formate
B: 50:50 Methanol: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
101000
242080
34.11000
471000
Flow Rate:
0.5 mL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 4000™

Order Items Used in This Application

Similar Applications
search
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase
Synergi 2.5 µm Hydro-RP 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase