FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Hi Temp Stability - Injection on Aeris WIDEPORE after 1050 cycles
19833
Separation Mode: Reversed Phase
Reversed Phase
WIDEPORE XB-C18
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Chromatographic Property
Biochemical Compound / Nutrient
Therapeutic / Clinical
Biochemical Compound / Nutrient
Biochemical Compound / Nutrient
Therapeutic / Clinical

Applications

Hi Temp Stability - Injection on Aeris WIDEPORE after 1050 cycles
Analytes
1Apomyoglobin

Analyte not found

Details
LC Conditions (App ID: 19833)
Column:
Brand Name:
Part No:
Phase Name:
WIDEPORE XB-C18
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% TFA
B: Acetonitrile with 0.1% TFA
Gradient Profile:
Step No.Time(min)%A%B
10973
23973
3233565
Flow Rate:
0.312 mL/min
Temperature:
90°C
Detection:
UV-Vis Abs.-Diode Array (PDA) (22°C)
Sample Note:
Test was run on a 50x4.6 Aeris WIDEPORE 3.6u XB-C18 run at 1.5 mL/min which is not commercially available at this time. Application was modifed to reflect the use of a 2.1 mm column for those that wish to replicat the data

Order Items Used in This Application

Similar Applications
search
Aeris 3.6 µm PEPTIDE XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C18 200 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C18 200 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna Omega 1.6µm PS C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Luna Omega 1.6µm Polar C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Luna Omega 1.6µm C18 100Å, LC Column 50 x 2.1 mm, EA
Reversed Phase
Aeris 3.6 µm PEPTIDE XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C18 200 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Aeris 3.6 µm WIDEPORE XB-C18 200 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase