For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Benzodiazepines using Kinetex 2.6u C18 50x2.1mm
19906
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Antiemetics
Clinical
Forensics/Toxicology

Applications

Benzodiazepines using Kinetex 2.6u C18 50x2.1mm
Analytes
1alpha-Hydroxyalprazolam
2Oxazepam
3Lorazepam
4Clonazepam
5Nordiazepam
6Temazepam
7Diazepam
Compound Name
Alpha-Hydroxyalprazolam
CID
162244
Molecular Formula
C17H13ClN4O
Molecular Weight
324
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 19906)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 10mM ammonium formate (no pH adjustment)
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
108515
244555
Flow Rate:
0.5 mL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 4000™

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase
Kinetex 2.6 µm Phenyl-Hexyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna 3 µm PFP(2) 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 3 mm, Ea
Reversed Phase
Kinetex 2.6 µm Phenyl-Hexyl 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase