For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Phthalates by GCMS on a ZB-50 30x0.25x0.25
19931
Separation Mode: Standard
Standard
ZB-50
GC
Teratogens
Food & Beverage

Applications

Phthalates by GCMS on a ZB-50 30x0.25x0.25
Analytes
1Dimethyl phthalate
2Diethyl phthalate
3Diallyl phthalate
4Dipropyl phthalate
5Dibutyl phthalate
6Diisobutyl phthalate
7Dihexyl phthalate
8bis(2-Methoxyethyl)phthalate
9Dipentyl phthalate
10bis(2-Ethoxyethyl)phthalate
11bis(4-Methyl-2-pentyl)phthalate
12Di(ethylhexyl) phthalate
13Butylbenzyl phthalate
14Diheptyl phthalate
15bis(2-Butoxyethyl)phthalate
16Dicyclohexyl phthalate
17Di-n-octyl phthalate
18Diphenylhexyl phthalate
19Dinonyl phthalate
20Diisononyl phthalate
21Diisodecyl phthalate
Compound Name
Dimethyl Phthalate
CID
8554
Molecular Formula
C10H10O4
Molecular Weight
194
No. Hydrogen Bond Acceptors
4
Details
GC Conditions (App ID: 19931)
Column:
Part No:
Phase Name:
ZB-50
Injection:
Split 10:1 @ 260°C, 1.0 μL
Oven Program:
135 ºC to 275 ºC @ 25 ºC/min for 3.5 min to 340 ºC @ 35 ºC/min for 1 min
Carrier Gas:
Constant Flow Helium, 1.0 mL/min
Detection:
Mass Selective (MSD) @ 45 amu (320°C)
Sample Notes:
Analytes are 100µg/mL

Order Items Used in This Application

Similar Applications
search
Kinetex 5 µm C18 100 Å, LC Column 100 x 4.6 mm, Ea
Reversed Phase
Luna 5 µm CN 100 Å, LC Column 150 x 4.6 mm, Ea
Normal Phase
Luna 5 µm CN 100 Å, LC Column 150 x 4.6 mm, Ea
Normal Phase
Zebron ZB-5, GC Cap. Column 30 m x 0.53 mm x 1.50 µm, Ea
Standard