FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Phthalates in Foods by LC/MS /MS on Synergi Polar-RP 2.5µm, 100x2.0 mm
19932
Separation Mode: Reversed Phase
Reversed Phase
Polar-RP
HPLC
Plasticizers

Applications

Phthalates in Foods by LC/MS /MS on Synergi Polar-RP 2.5µm, 100x2.0 mm
Analytes
1bis(2-Methoxyethyl)phthalate
2Dimethyl phthalate
3bis(2-Ethoxyethyl)phthalate
4Diethyl phthalate
5Diallyl phthalate
6Di-n-propyl phthalate
7Diisobutyl phthalate
8Di-n-butyl phthalate
9bis(2-n-Butoxyethyl)phthalate
10Dibenzylbutyl phthalate
11Dipentyl phthalate
12Dicyclohexyl phthalate
13Di-4-methyl-2-pentyl phthalate
14Dihexyl phthalate
15Diheptyl phthalate
16bis(2-Ethylhexyl)phthalate
17Di-n-octyl phthalate
18Diisononyl phthalate
19Di-n-nonyl phthalate
20Diisodecyl phthalate
Compound Name
Kesscoflex Mcp
CID
8344
Molecular Formula
C14H18O6
Molecular Weight
282
No. Hydrogen Bond Acceptors
6
Details
LC Conditions (App ID: 19932)
Column:
Brand Name:
Part No:
Phase Name:
Polar-RP
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid in water
B: 0.1% Formic acid in 50:50 Acetonitrile: Methanol
Gradient Profile:
Step No.Time(min)%A%B
106040
215050
365050
4111585
5141585
614.16040
7196040
Flow Rate:
0.4 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 4000™
Sample Note:
SecurityGuard™ ULTRA Guard Cartridge System extends column lifetime. - SecurityGuard ULTRA Cartridges, UHPLC Phenyl for 2.1mm ID Columns, 3/Pk Part No.: AJ0-8788 - Holder Part No.: AJ0-9000

Order Items Used in This Application