For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Pharmaceutical Drug Mix on Kinetex C18 2.6u 75 x 2.1mm @ 1.1 mL/min(Quotient Data)
19939
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Antidepressive Agents, Tricyclic
Pharmaceutical/Biopharmaceutical

Applications

Pharmaceutical Drug Mix on Kinetex C18 2.6u 75 x 2.1mm @ 1.1 mL/min(Quotient Data)
Analytes
1Haloperidol
2Diltiazem
3Terfenadine
4Cimetidine
5Acetaminophen
6Sulfathiazole
7Pindolol
8Quinidine
9Acebutolol
10Chlorpheniramine
11Triprolidine
12Prednisolone
13Nortriptyline
142-Hydroxy-5-methylbenzaldehyde
15Hexanophenone
Compound Name
Haloperidol
CID
3559
Molecular Formula
C21H23ClFNO2
Molecular Weight
375
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 19939)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% formic acid in acetonitrile
B: 0.1% formic acid
Gradient Profile:
Step No.Time(min)%A%B
10595
20.25595
31.89010
41.99010
51.91595
62.5595
Flow Rate:
1.1 mL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)