For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Separation of Benzodiazepines
20047
Separation Mode: Reversed Phase
Reversed Phase
Hydro-RP
HPLC
Antiemetics
Clinical
Forensics/Toxicology
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Separation of Benzodiazepines
Analytes
17-Aminoclonazepam
27-Aminoflurazepam
3Midazolam
4Flurazepam
5a-OH-Alprazolam
6a-OH-Triazolam
7Oxazepam
8Lorazepam
9Clonazepam
10Nordiazepam
11Alprazolam
12Flunitrazepam
13Temazepam
14Diazepam
Compound Name
7-Aminoclonazepam
CID
188298
Molecular Formula
C15H12ClN3O
Molecular Weight
285
No. Hydrogen Bond Acceptors
3
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 20047)
Column:
Brand Name:
Part No:
Phase Name:
Hydro-RP
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid in DI H2O
B: 0.1% formic Acid in ACN
Gradient Profile:
Step No.Time(min)%A%B
10.59010
20.78020
35.92575
46595
56.2595
66.39010
78.59010
Flow Rate:
500 µL/min
Temperature:
40°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
3200 QTRAP/API 3200
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, AQ C18 4 x 2.0mm ID, 10/Pk Part No.: AJ0-7510 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 30 x 2.1 mm, Ea
Reversed Phase
Luna 5 µm C18(2) 100 Å, LC Column 50 x 2 mm, Ea
Reversed Phase