For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Human Plasma Vitamin C on Kinetex 5u XB-C18, 150x4.6mm Wavelength Study
21442
Separation Mode: Reversed Phase
Reversed Phase
XB-C18
HPLC
Antioxidants
Clinical

Applications

Human Plasma Vitamin C on Kinetex 5u XB-C18, 150x4.6mm Wavelength Study
Analytes
1Vitamin C
Compound Name
Ascorbic Acid
CID
5785
Molecular Formula
C6H8O6
Molecular Weight
176
No. Hydrogen Bond Acceptors
6
No. Hydrogen Bond Donors
4
Details
LC Conditions (App ID: 21442)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
XB-C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% formic acid
B: Acetonitrile 100%
Gradient Profile:
Step No.Time(min)%A%B
101000
23.51000
33.60100
450100
55.11000
671000
Flow Rate:
800 µL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 245 nm (22°C)
Sample Note:
Guard column and cartridges were used for human plasma vitamin C analysis: SecurityGuard Ultra Cartridges: Cat log No. AJO-8768. SecurityGuard Ultra Cartridge holder: Cat log No. AJO-9000.
Sample Preparation Details
Description:
Impact Protein Precipitation, 2mL Square Well Filter Plate, 2/Pk
Part Number:
Evaporation:
Evaporate to dryness

Order Items Used in This Application

Similar Applications
search
Kinetex 5 µm XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase
Kinetex 5 µm XB-C18 100 Å, LC Column 150 x 4.6 mm, Ea
Reversed Phase