For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Extraction of Total Cortisol from Human Plasma Using Novum SLE & Kinetex Biphenyl
22787
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Anti-Inflammatory Agents
Clinical
Clinical Research
Therapeutic / Clinical

Applications

Extraction of Total Cortisol from Human Plasma Using Novum SLE & Kinetex Biphenyl
Analytes
1Cortisol
2Cortisol-D4 (Internal standard)
Compound Name
Hydrocortisone
CID
5754
Molecular Formula
C21H30O5
Molecular Weight
362
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
3
Details
LC Conditions (App ID: 22787)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Biphenyl
Elution Type:
Gradient
Mobile Phase:
A: 10 mm ammonium acetate in water
B: 10 mm ammonium acetate in methanol
Gradient Profile:
Step No.Time(min)%A%B
105050
22595
33.1595
43.115050
555050
Flow Rate:
0.45 mL/min
Temperature:
25°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX API 5000
Sample Preparation Details
Description:
Novum SLE, MAX 96-Well Plate, Ea
Part Number:
Pre Treatment Note:
Extraction Procedure Sample pre-treatment: Dilute 200 uL of human plasma with 200 ul of 50 mM sodium phosphate dibasic heptahydrate, pH unadjusted. Vortex briefly (3-5 sec). Sample loading: Load the sample from pretreatment step abive onto the Novum plate and apply a short and gentle pulse of vacuum (~ 5" of Hg for 5-10 sec) or until the sample has completely entered the media. Wait for 5 minutes. Elution: Dispense 900 uL of dichloromethane onto the SLE media and allow the solvent to elute by gravity (~ 5 min elution time) and collect the eluant. Repeat with another 900 uL of dichloromethane and collect the eluant. Apply vacuum at 5" of Hg for 20-30 secs to complete the extraction. NOTE: Prolonged application of vacuum will result in elution of plasma out of the SLE media and into the final extracted solvent. Dry down: Evaporate the final extract to complete dryness under slow stream of N2 at 40°C Reconstitute the dry residue in 200 uL of initial mobile phase fortified with cortisol-D4.
Sample PreTreatment:
Extraction Procedure Sample pre-treatment: Dilute 200 uL of human plasma with 200 ul of 50 mM sodium phosphate dibasic heptahydrate, pH unadjusted. Vortex briefly (3-5 sec). Sample loading: Load the sample from pretreatment step abive onto the Novum plate and apply a short and gentle pulse of vacuum (~ 5" of Hg for 5-10 sec) or until the sample has completely entered the media. Wait for 5 minutes. Elution: Dispense 900 uL of dichloromethane onto the SLE media and allow the solvent to elute by gravity (~ 5 min elution time) and collect the eluant. Repeat with another 900 uL of dichloromethane and collect the eluant. Apply vacuum at 5" of Hg for 20-30 secs to complete the extraction. NOTE: Prolonged application of vacuum will result in elution of plasma out of the SLE media and into the final extracted solvent. Dry down: Evaporate the final extract to complete dryness under slow stream of N2 at 40°C Reconstitute the dry residue in 200 uL of initial mobile phase fortified with cortisol-D4.
Evaporation:
Evaporate to dryness
1
Shake
Methylene Chloride
2
Shake
Methylene Chloride

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm Biphenyl 100 Å, LC Column 50 x 3.0 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 100 x 3 mm, Ea
Reversed Phase