For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
FAMEs In Soybean Oil by GC/FID on Zebron ZB-FAME (30 m)
23860
Separation Mode: Application Specific
Application Specific
ZB-FAME
GC
Enzyme Inhibitors
Food & Beverage

Applications

FAMEs In Soybean Oil by GC/FID on Zebron ZB-FAME (30 m)
Analytes
1C11:0
2C16:0
3C18:0
4C18:1 cis 9
5C18:2 cis 9,12
6C18:3 cis 9,12,15
7C20:0
8C20:1 cis 11
9C22:0
Compound Name
Methyl Undecanoate
CID
15607
Molecular Formula
C12H24O2
Molecular Weight
200
No. Hydrogen Bond Acceptors
2
Details
GC Conditions (App ID: 23860)
Column:
Part No:
Phase Name:
ZB-FAME
Injection:
Split 50:1 @ 240°C, 1.0 μL
Oven Program:
100 °C for 2 min to 140 °C @ 10 °C/min to 190 °C @ 3 °C/min to 260 °C @ 30 °C/min for 2 min
Carrier Gas:
Constant Flow Helium, 1.2 mL/min
Detection:
Flame Ionization (FID) (260°C)
Sample Notes:
Recommended Liner: Zebron PLUS Single Taper with Wool, 4 mm ID Liner Part No.: AG2-0A11-05 (for Agilent systems) Inlet Seal: AG0-8620 (Gold Plated Easy Seal) Septum: AG0-4696 (PhenoRed-400) Sample Preparation: 1. Strata® Si-1 Tube, 1 g/6 mL (Part No.: 8B-S012-JCH) on a vacuum or positive pressure manifold 2. Wash cartridge with 6 mL of hexane 3. Load oil solution (0.12 g of oil in 0.5 mL of hexane) 4. Elute with 10 mL of hexane/diethyl ether (87:13) 5. Evaporate eluate under a steady stream of nitrogen 6. Dissolve purified oil residue in 1mL of heptane 7. Add 0.1 mL of 2N potassium hydroxide in methanol to purified oil solution 8. Cap tube and shake vigorously for 15 seconds 9. Leave to separate until upper layer becomes clear 10. Extract upper layer for GC analysis
Sample Preparation Details
Description:
Strata SI-1 Silica (55 µm, 70 Å), 1 g / 6 mL, Tubes, 30/Pk
Part Number:
Pre Treatment Note:
To reconstituted sample, add 0.1 mL of 2 N Potassium hydroxide in methanol. Cap tube and shake vigorously for 15 seconds. Leave to separate until upper layer becomes clear. Extract upper layer for GC anlaysis.
Condition 1:
Hexane
Sample PreTreatment:
To reconstituted sample, add 0.1 mL of 2 N Potassium hydroxide in methanol. Cap tube and shake vigorously for 15 seconds. Leave to separate until upper layer becomes clear. Extract upper layer for GC anlaysis.
Evaporation:
Evaporate to dryness
1
Condition
Hexane
1
Shake
Hexane/Diethyl ether (87:13)

Order Items Used in This Application

Similar Applications
search
No Data