Mercury - Progesterones
LC Conditions
Column
Luna 3 µm C18(2) 100 Å, LC Column 20 x 2 mm, Ea
Brand
Luna
Part No
00M-4251-B0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: Water with 0.1% formic acid
B: Acetonitrile with 0.1% formic acid
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
3
5
95
Flow Rate
0.80 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 2.0mm, 10/Pk Part No.: AJ0-4286 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. 11-alpha-Hydroxyprogesterone
- 2. 17-alpha-Hydroxyprogesterone
- 3. Progesterone
Similar Applications
No Similar applications found.