Diisocyanate
LC Conditions
Column
Luna 5 µm Phenyl-Hexyl 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Luna
Part No
00G-4257-E0
Phase Name
Phenyl-Hexyl
Elution Type
Gradient
Mobile Phase
A: 0.007M Ammonium acetate, pH 6.0/Acetonitrile (67.5:32.5 v/v)
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
100
0
2
5
100
0
3
20
75
25
4
30
65
35
5
35
100
0
6
40
100
0
Flow Rate
0.60 mL/min
Temperature
20°C
Detection
Fluorescence (FLUOR) @ (22°C)
Detection Info
Sample Notes
The derivatisation reagent used was 1-(2-pyridyl)piperazine (1-2PP). SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, Phenyl 4 x 3.0mm, 10/Pk Part No.: AJ0-4351 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. HDI
- 2. 2,4-TDI
- 3. MDI
Similar Applications
No Similar applications found.