Nicotine and Related Compounds
LC Conditions
Column
Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Brand
Synergi
Part No
00F-4337-E0
Phase Name
Max-RP
Elution Type
Gradient
Mobile Phase
A: 10 mM Ammonium acetate in water
B: 10 mM Ammonium acetate in methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
90
10
2
0.25
90
10
3
3.75
0
100
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C12 4 x 3.0mm, 10/Pk Part No.: AJ0-6074 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Cotinine-N'-oxide
- 2. 1'(R)-2'(S)-cis-Nicotine-N'-oxide
- 3. 1'(S)-2'(S)-trans-Nicotine-N'-oxide
- 4. Hydroxycotinine
- 5. Cotinine
- 6. Nicotine
- 7. Picric acid
Similar Applications
No Similar applications found.