Corticosteroids
LC Conditions
Column
Luna 3 µm CN 100 Å, LC Column 100 x 4.6 mm, Ea
Brand
Luna
Part No
00D-4254-E0
Phase Name
CN
Elution Type
Gradient
Mobile Phase
A: Hexane
B: 80:20 Methylene chloride:Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
2
70
30
Flow Rate
4.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, CN 4 x 3.0mm, 10/Pk Part No.: AJ0-4305 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Desoxycorticosterone
- 2. 17-alpha-Hydroxyprogesterone
- 3. alpha-Estradiol
- 4. 17-alpha-Estradiol
- 5. Cortisone acetate
- 6. Hydroxycortisone acetate
- 7. Corticosterone
- 8. Cortisone
- 9. Prednisone
- 10. Hydrocortisone
- 11. Prednisolone
Similar Applications