Narcotic Analgesics
LC Conditions
Column
Synergi 4 µm Polar-RP 80 Å, LC Column 150 x 2 mm, Ea
Brand
Synergi
Part No
00F-4336-B0
Phase Name
Polar-RP
Elution Type
Gradient
Mobile Phase
A: 95% Methanol / 5% Water w/0.05% Formic acid
B: 5% Methanol / 95% Water w/0.05% Formic acid
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
20
80
2
30
90
10
3
35
90
10
4
38
20
80
Flow Rate
0.20 mL/min
Temperature
12°C
Detection
Mass Spectrometer (MS) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, POLAR-RP 4 x 2.0mm, 10/Pk Part No.: AJ0-6075 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Dihydrocodeine
- 2. Codeine-d3
- 3. Hydrocodone-d6
Similar Applications