EPA Method 8330 - Explosives on Luna C18(2)
LC Conditions
Official Method
EPA 8330
Column
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Luna
Part No
00G-4252-E0
Phase Name
C18(2)
Elution Type
Isocratic
Mobile Phase
A: Methanol
B: Water
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
55
45
Flow Rate
1.00 mL/min
Temperature
35°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. HMX
- 2. RDX
- 3. 1,3,5-Trinitrobenzene
- 4. 1,3-Dinitrobenzene
- 5. Tetryl
- 6. Nitrobenzene
- 7. 2,4,6-Trinitrotoluene
- 8. 4-Amino-2,6-dinitrotoluene
- 9. 2-Amino-4,6-dinitrotoluene
- 10. 2,6-Dinitrotoluene
- 11. 2,4-Dinitrotoluene
- 12. 2-Nitrotoluene
- 13. 4-Nitrotoluene
- 14. 3-Nitrotoluene