Extraction of Bronchodilators on Strata™ X

Extraction of Bronchodilators on Strata™ X
LC Conditions
Column

Luna 5 µm C8(2) 100 Å, LC Column 150 x 4.6 mm, Ea

Brand

Luna

Part No

00F-4249-E0

Phase Name

C8(2)

Elution Type

Isocratic

Mobile Phase

A: 20 mM KH2PO4, pH 2.5

B: ACN

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

95

5

Flow Rate

1.50 mL/min

Temperature

25°C

Detection

UV-Vis Abs.-Fixed Wave. (UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C8 4 x 3.0mm, 10/Pk Part No.: AJ0-4290 - Holder Part No.: KJ0-4282

Sample Preparation Details
Description

Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk

Part No

8B-S100-TAK

Pretreatment Note

Blow down to almost dryness under slow stream of N2. Reconstitute with mobile phase to 200 uL

Condition 1

1.0 mL Methanol

Condition 2

1.0 mL Water

Load

Load 1.00 mL Sample diluted in Buffer

Wash 1

1.0 mL 5% methanol/D. I. water

Wash 2

1 at 2

Dry

for 1 min at 10'' of Hg

Elution 1

1.0 mL Methanol

Evaporation

Evaporate to dryness

Reconstitute

2 mL Mobile Phase

Order Items Used in This Application

Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
8B-S100-TAK

Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk

View Product

Luna 5 µm C8(2) 100 Å, LC Column 150 x 4.6 mm, Ea
00F-4249-E0

Luna 5 µm C8(2) 100 Å, LC Column 150 x 4.6 mm, Ea

View Product

  • 1. Theobromine
  • 2. Theophylline
  • 3. Beta hydroxy theophylline
  • 4. Caffeine
-OEChem-06211003152D<br><br> 21 22  0     0  0  0  0  0  0999 V2000<br>    3.7320    2.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.0000   -1.0000    0.0000 O   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.5443    0.8047    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -1.0000    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660    0.5000    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.5443   -0.8047    0.0000 N   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981    0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.5981   -0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320    1.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.8660   -0.5000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.1279    0.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -2.0000    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.8550    1.7552    0.0000 C   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.7479    0.0000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    2.3291    0.8100    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.1120   -2.0000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    3.7320   -2.6200    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    4.3520   -2.0000    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.4443    1.5626    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    6.0476    2.3446    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>    5.2656    1.9479    0.0000 H   0  0  0  0  0  0  0  0  0  0  0  0<br>  1  9  2  0  0  0  0<br>  2 10  2  0  0  0  0<br>  3  7  1  0  0  0  0<br>  3 11  1  0  0  0  0<br>  3 13  1  0  0  0  0<br>  4  8  1  0  0  0  0<br>  4 10  1  0  0  0  0<br>  4 12  1  0  0  0  0<br>  5  9  1  0  0  0  0<br>  5 10  1  0  0  0  0<br>  5 15  1  0  0  0  0<br>  6  8  1  0  0  0  0<br>  6 11  2  0  0  0  0<br>  7  8  2  0  0  0  0<br>  7  9  1  0  0  0  0<br> 11 14  1  0  0  0  0<br> 12 16  1  0  0  0  0<br> 12 17  1  0  0  0  0<br> 12 18  1  0  0  0  0<br> 13 19  1  0  0  0  0<br> 13 20  1  0  0  0  0<br> 13 21  1  0  0  0  0<br>M  END
Compound Name

Theobromine

CID

5429

Molecular Formula

C7H8N4O2

Molecular Weight

180

No. Hydrogen Bond Acceptors

3

No. Hydrogen Bond Donors

1

Similar Applications

Reversed Phase

Caffeine & Metabolites, pH 7.5 - Luna 5µm C18(2)

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

15662

Reversed Phase

Caffeine & Metabolites, pH 7.5 (Synergi 4µm Fusion-RP)

Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea

15661

Reversed Phase

Caffeine & Metabolites, pH 10.0 - Luna 5µm C18(2)

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

15660

Reversed Phase

Caffeine & Metabolites, pH 10.0 (Synergi 4µm Fusion-RP)

Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea

15659

Reversed Phase

Caffeine & Metabolites, pH 4.5 (Synergi 4µm Fusion-RP)

Synergi 4 µm Fusion-RP 80 Å, LC Column 150 x 4.6 mm, Ea

15658

Reversed Phase

Caffeine & Metabolites, pH 4.5 - Luna 5µm C18(2)

Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea

15657

Reversed Phase

Caffeine & Metabolites, pH 7.5

Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea

15014

Reversed Phase

Caffeine & Metabolites, pH 4.5

Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea

15013

Reversed Phase

Caffeine & Metabolites, pH 10.0

Gemini 5 µm C18 110 Å, LC Column 150 x 4.6 mm, Ea

15012

Reversed Phase

Green Tea

Luna 5 µm Phenyl-Hexyl 100 Å, LC Column 250 x 4.6 mm, Ea

14436

Reversed Phase

Caffeine and related compounds - Synergi Max-RP

Synergi 4 µm Max-RP 80 Å, LC Column 150 x 4.6 mm, Ea

14213