EPA Method 610-QC-FL - PAH's on Luna 5um C18(2)

EPA Method 610-QC-FL - PAH's on Luna 5um C18(2)
LC Conditions
Official Method

EPA 610

Column

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea

Brand

Luna

Part No

00G-4252-E0

Phase Name

C18(2)

Elution Type

Gradient

Mobile Phase

A: Water

B: Acetonitrile

Gradient Profile

Step No.

Time(min)

%A

%B

1

0

60

40

2

5

60

40

3

30

0

100

Flow Rate

1.55 mL/min

Temperature

22°C

Detection

UV-Vis Abs.-Variable Wave.(UV) @ (22°C)

Detection Info
Sample Notes

SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
00G-4252-E0

Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea

View Product

  • 1. Naphthalene
  • 2. Acenaphthylene
  • 3. 1-Methylnaphthalene
  • 4. 2-Methylnaphthalene
  • 5. Acenaphthene
  • 6. Fluorene
  • 7. Phenanthrene
  • 8. Anthracene
  • 9. Fluoranthene
  • 10. Pyrene
  • 11. Benz[a]anthracene
  • 12. Chrysene
  • 13. Benzo[b]fluoranthene
  • 14. Benzo[k]fluoranthene
  • 15. Benzo[a]pyrene
  • 16. Dibenz[a,h]anthracene
  • 17. Benzo[g,h,i]perylene
  • 18. Indeno[1,2,3-cd]pyrene
Similar Applications