EPA Method 553, Benzidines and Nitrogen-Containing Pesticides
LC Conditions
Official Method
EPA 553
Column
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna
Part No
00F-4252-E0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: 10mM Ammonium Acetate:Acetonitrile (75:25)
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
75
25
2
1
75
25
3
29
40
60
4
49
40
60
Flow Rate
0.30 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282