EPA Method 610, PAH's
LC Conditions
Official Method
EPA 610
Column
Synergi 4 µm Max-RP 80 Å, LC Column 250 x 4.6 mm, Ea
Brand
Synergi
Part No
00G-4337-E0
Phase Name
Max-RP
Elution Type
Gradient
Mobile Phase
A: Water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
60
40
2
5
60
40
3
30
0
100
Flow Rate
1.55 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C12 4 x 3.0mm, 10/Pk Part No.: AJ0-6074 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Naphthalene
- 2. Acenaphthylene
- 3. Acenaphthene
- 4. Fluorene
- 5. Phenanthrene
- 6. Anthracene
- 7. Fluoranthene
- 8. Pyrene
- 9. Benz[a]anthracene
- 10. Chrysene
- 11. Benzo[b]fluoranthene
- 12. Benzo[k]fluoranthene
- 13. Benzo[a]pyrene
- 14. Dibenz[a,h]anthracene
- 15. Benzo[g,h,i]perylene
- 16. Indeno[1,2,3-cd]pyrene
Similar Applications