EPA Method 632.1-632.4, Carbamate and Amide Pesticides
LC Conditions
Official Method
EPA 632
Column
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6 mm, Ea
Brand
Luna
Part No
00F-4252-E0
Phase Name
C18(2)
Elution Type
Isocratic
Mobile Phase
A: Water
B: Acetonitrile
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
45
55
Flow Rate
2.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Vacor
- 2. Carbaryl
- 3. Propanil
- 4. Napropamide
Similar Applications