EPA Method 8318, N-Methylcarbamates
LC Conditions
Official Method
EPA 8318
Column
Luna 5 µm C18(2) 100 Å, LC Column 250 x 4.6 mm, Ea
Brand
Luna
Part No
00G-4252-E0
Phase Name
C18(2)
Elution Type
Gradient
Mobile Phase
A: 0.4% Phosphoric Acid
B: 50:50 Acetonitrile:Methanol
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
80
20
2
20
0
100
3
30
0
100
Flow Rate
1.00 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282
Order Items Used in This Application
- 1. Aldicarb
- 2. Methomyl
- 3. 3-Hydroxycarbofuran
- 4. Dioxacarb
- 5. Aldicarb
- 6. Propoxur
- 7. Carbofuran
- 8. Carbaryl
- 9. Methiocarb
- 10. Promecarb
Similar Applications
No Similar applications found.