Peptides from Serum on Strata™ X
LC Conditions
Column
Jupiter 4 µm Proteo 90 Å, LC Column 150 x 4.6 mm, Ea
Brand
Jupiter
Part No
00F-4396-E0
Phase Name
Proteo
Elution Type
Gradient
Mobile Phase
A: 0.1% TFA
B: ACN:0.1% TFA in water (90:10)
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
85
15
2
22
40
60
Flow Rate
1.00 mL/min
Temperature
30°C
Detection
UV-Vis Abs.-Diode Array (PDA) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C12 4 x 3.0mm, 10/Pk Part No.: AJ0-6074 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Pretreatment Note
EVAPORATE AND RECONSTITUTE .40 °C under nitrogen stream 300 mL 0.1% aqueous TFA
Condition 1
1.0 mL Methanol
Condition 2
1.0 mL Water
Sample Pretreatment
Strata X 33u 30 mg / 1 ml Tube Condition: 1ml Methanol Load: 1 ml Spiked porcine serum extract
Load
Load 1.00 mL Sample
Wash 1
1.0 mL 10% Acetonitrile in 0.1% TFA
Dry
1 min @ 10 in. Hg
Elution 1
1.0 mL Acetonitrile:Water
Evaporation
Evaporate to dryness
Reconstitute
300 µL TFA (0.1%) in Water
Order Items Used in This Application
- 1. Oxytocin
- 2. Angiotensin II
- 3. Angiotensin I
Similar Applications