Extraction of Diazepam on Strata™ X
LC Conditions
Column
Synergi 4 µm Polar-RP 80 Å, LC Column 150 x 4.6 mm, Ea
Brand
Synergi
Part No
00F-4336-E0
Phase Name
Polar-RP
Elution Type
Gradient
Mobile Phase
A: 20 mM KH2PO4 (pH=2.5)
B: MeOH
Gradient Profile
Step No.
Time(min)
%A
%B
1
0
75
25
2
3
75
25
3
8
30
70
4
11
30
70
5
14
75
25
Flow Rate
1.50 mL/min
Temperature
22°C
Detection
UV-Vis Abs.-Variable Wave.(UV) @ (22°C)
Detection Info
Sample Notes
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, POLAR-RP 4 x 3.0mm, 10/Pk Part No.: AJ0-6076 - Holder Part No.: KJ0-4282
Sample Preparation Details
Description
Strata-X 33 µm Polymeric Reversed Phase, 30 mg / 1 mL, Tubes, 100/Pk
Part No
8B-S100-TAK
Pretreatment Note
Reconstitute with 200 µL mobile phase
Condition 1
1.0 mL Methanol
Condition 2
1.0 mL Water
Load
Load 1000.00 µL Sample
Wash 1
1.0 mL Methanol:Water
Dry
1 min @ 10 in. Hg
Elution 1
Methanol
Evaporation
Evaporate to dryness
Reconstitute
200 µL Mobile Phase
Order Items Used in This Application
- 1. Desmethyldiazepam
- 2. Diazepam
Similar Applications
No Similar applications found.