Polyaromatichydrocarbons (PAH):EPA Method 610 - ZB-5 30x0.25x0.25

Polyaromatichydrocarbons (PAH):EPA Method 610 - ZB-5 30x0.25x0.25
GC Conditions
Official Method

EPA 610

Column

Zebron ZB-5, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

Part No

7HG-G002-11

Phase Name

ZB-5

Injection

Split 15:1 @ 260°C

Carrier Gas

Constant Flow Helium, 1.4 mL/min

Detection

Mass Selective (MSD) @ (240°C)

Detection Info

Order Items Used in This Application

Zebron ZB-5, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
7HG-G002-11

Zebron ZB-5, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea

View Product

  • 1. Naphthalene
  • 2. 2-Methylnaphthalene
  • 3. Acenaphthylene
  • 4. Acenaphthene
  • 5. Fluorene
  • 6. Phenanthrene
  • 7. Anthracene
  • 8. Fluoranthene
  • 9. Pyrene
  • 10. Benz[a]anthracene
  • 11. Chrysene
  • 12. Benzo[b]fluroanthene
  • 13. Benzo[k]fluroanthene
  • 14. Benzo[a]pyrene
  • 15. Indeno[1,2,3-cd]pyrene
  • 16. Dibenz[a,h]anthracene
  • 17. Benzo[g,h,i]perylene
Similar Applications